4.3.1 Main Tab

Communication and utilities are accessed on the Main tab. (Refer to

Figure 4-2

.)

NOTE: Right-clicking on a value or parameter and then clicking Description and

Tip will display additional information about that value or parameter.

4.3.1.1 Main Setup Pane

Communication Settings

Com Port# . . . . . . . . . . . . . . . . . . . . 1 to 255

Sets the communications port that STM-2XM
is communicating through.

Protocol. . . . . . . . . . . . . . . . . . . . . . Sycon @ 9.6 Kbps, SMDP L (9.6 Kbps),

SMDP M (38.4 Kbps), SMDP H (115.2 Kbps).

Adjustable baud rate.

NOTE: This value must match the Comm

Protocol setting in the System
Settings menu on STM-2XM (refer to

section 3.2.1.2 on page 3-10

).

Slave Address. . . . . . . . . . . . . . . . . 16 to 255

SMDP address the software will use for serial
communication with STM-2XM.

NOTE: This value must match the SMDP

Address setting in the System
Settings menu on STM-2XM (refer to

section 3.2.1.2 on page 3-10

).

Timeout(ms) . . . . . . . . . . . . . . . . . . 150 to 20000

The amount of time in microseconds that the
software will wait for a response before
issuing an error.

Program Control

Exit Program . . . . . . . . . . . . . . . . . . Click to safely stop communications and exit

the software program.

Purge/Re-Init all edit controls . . . . Returns all parameters edited using the

software to default values.