4 - 9

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

Paste film . . . . . . . . . . . . . . . . . . . . Click to paste the film parameters for the

displayed film in the film on clipboard box
onto the current film parameters displayed,
thereby replacing the values for the current
film.

All films pane

Use the spin box to increment or decrement the values for each of the parameters
listed below.

EDITING film# . . . . . . . . . . . . . . . . . 1 to 15

Name . . . . . . . . . . . . . . . . . . . . . . . . 8 character maximum

Density . . . . . . . . . . . . . . . . . . . . . . 0.40 to 99.99 g/cm

3

Z-Ratio . . . . . . . . . . . . . . . . . . . . . . . 0.100 to 9.999

Tooling. . . . . . . . . . . . . . . . . . . . . . . 10 to 999.9%

Setpoint mass. . . . . . . . . . . . . . . . . 0.00 to 9999.99 µg/cm²

Setpoint Thickness . . . . . . . . . . . . 0.000 to 999.999 kÅ

End mass . . . . . . . . . . . . . . . . . . . . 0.00 to 9999.99 µg/cm²

End Thickness . . . . . . . . . . . . . . . . 0.000 to 999.999 kÅ

Use Backup. . . . . . . . . . . . . . . . . . . Select to enable the use of the backup crystal

(refer to

section 3.2.1.1 on page 3-8

).

Setpoint Time . . . . . . . . . . . . . . . . . 00:00:00 to 9:59:59 h:mm:ss

Desired Rate . . . . . . . . . . . . . . . . . . 0.00 to 99.99 µg•s/cm²

Desired Rate . . . . . . . . . . . . . . . . . . 0.0 to 999.9 Å/s

Do shutter delay? . . . . . . . . . . . . . . Select to enable shutter delay (refer to

section 3.2.1.1 on page 3-8

).

Do rate sampling . . . . . . . . . . . . . . Select to enable rate sampling (refer to

section 3.2.1.1 on page 3-8

).

Sampling interval . . . . . . . . . . . . . . 00:00:00 to 9:59:59 h:mm:ss

Sampling Dwell. . . . . . . . . . . . . . . . 00:00:00 to 9:59:59 h:mm:ss

Do time/power. . . . . . . . . . . . . . . . . Select to enable time/power upon crystal fail

(refer to

section 3.2.1.1 on page 3-8

).

Xtal min life . . . . . . . . . . . . . . . . . . . 0 to 99%

Force fail on min life . . . . . . . . . . . Select to enable an automatic crystal fail

upon reaching the minimum crystal life input
(refer to

section 3.2.1.1 on page 3-8

).

# Samples . . . . . . . . . . . . . . . . . . . . 1 to 50

Good Rate Tolerance . . . . . . . . . . . 0 to 50%