2 system parameters pane – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 76

Advertising
background image

4 - 16

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

4.3.3.2 System parameters Pane

Main system settings
System Mode
. . . . . . . . . . . . . . . . . Sets STM-2XM to operate in a certain mode

(refer to

section 3.3 on page 3-17

). (See

Table 4-6

.

)

Test Mode . . . . . . . . . . . . . . . . . . . . Select to enable Test mode where sensor

readings are simulated.
NOTE: If STM-2XM is reset, test mode is

turned off automatically.

Active Channel/Source . . . . . . . . . CH1, CH2

Selects where the analog and digital outputs
and inputs are connected.

Display Mode . . . . . . . . . . . . . . . . . Displays thickness in kiloangstroms (kÅ) or

mass in micrograms per square centimeter
(µg/cm

3

).

Chart settings
Plot type
. . . . . . . . . . . . . . . . . . . . . Available in single film modes.

Displays the graph on the Runtime screen as
No plot, Rate Deviation, Rate, or
Thickness/Mass.

Plot time scale . . . . . . . . . . . . . . . . Sets the horizontal axis size. Defines the

amount of time covered by the plot before it
is overwritten.

Screen + sound
LCD Contrast
. . . . . . . . . . . . . . . . . 10 to 85

Adjusts the appearance of the screen.

LCD Brightness . . . . . . . . . . . . . . . 1 to 99

Adjusts the appearance of the screen.

Beeper on? . . . . . . . . . . . . . . . . . . . Select to enable a sound when a front panel

button is pressed.

Table 4-6 System mode

Mode

Action

Mode

Action

0

Simple, Sensor X1 active

3

Averaging

1

Simple, Sensor X2 active

4

Alloying

2

Sensor backup

5

Independent

Advertising