PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

Analog output 1 to 4 function . . . . Determines the function of each of the

analog outputs (refer to

section 2.2.3 on

page 2-5

for details on wiring analog

outputs). (See

Table 4-3

.)

Table 4-3 Analog output functions

Setting

Description

0: Zero volts

Output is disabled

1: +10 V Fixed

+10 V Fixed

2: -10 V Fixed

-10 V Fixed

3: Computer controlled

Output is computer set value

4: Rate Deviation, Channel 1, 1000 Å/S

Rate Deviation, Channel 1, 1000 Å/s

5: Rate Deviation, Channel 1, 10 Å/S

Rate Deviation, Channel 1, 10 Å/s

6: Rate, Channel 1, 1000 Å/S

Rate, Channel 1, 1000 Å/s

7: Rate, Channel 1, 10 Å/S

Rate, Channel 1, 10 Å/s

8: Thickness, Channel 1, 999 Å

Thickness Recorder, Channel 1, 999 Å

9: Thickness, Channel 1, 99.9 Å

Thickness Recorder, Channel 1, 99.9 Å

10: Rate Deviation, Channel 2, 1000 Å/S

Rate Deviation, Channel 2, 1000 Å/s

11: Rate Deviation, Channel 2, 10 Å/S

Rate Deviation, Channel 2, 10 Å/s

12: Rate, Channel 2, 1000 Å/S

Rate, Channel 2, 1000 Å/s

13: Rate, Channel 2, 10 Å/S

Rate, Channel 2, 10 Å/s

14: Thickness, Channel 2, 999 Å

Thickness Recorder, Channel 2, 999 Å

15: Thickness, Channel 2, 99.9 Å

Thickness Recorder, Channel 2, 99.9 Е