STM-2XM Operating Manual

5.8.11 Command: Q

Command Type . . . . . . . . . . . . . . . . Status

Command Structure . . . . . . . . . . . . . QTWIN_class, Indx X

Syntax . . . . . . . . . . . . . . . . . . . . . . . Qn,X

Description . . . . . . . . . . . . . . . . . . . . Displays Diagnostic information, where n is 1

to 21 and X is an index with Channel 1 as 0
and Channel 2 as 1.

Parameters. . . . . . . . . . . . . . . . . . . . (See

Table 5-7

.)

Table 5-7 Q command

#

Command Structure

Comments

Values

1 QTWIN_CH_T_MASS Channel

total

mass

µg

2 QTWIN_CH_T_THIK

Channel total thickness

3 QTWIN_XRAWF1

Raw frequencies both
channels

Hz

4 QTWIN_XAVGF1

Last known good averaged
frequencies

Hz

5 QTWIN_XLIFE

Last known good crystal life %

6 QTWIN_XRATEM

Last known good mass rate µg/cm²s

7 QTWIN_XRATET

Last known good thickness
rate

Å/s

8 QTWIN_XMASS

Accumulated mass

µg/cm²

9 QTWIN_XTHIK Accumulated

thickness

Å

10 QTWIN_XSTAT

Raw status (XtlStat_class)

Value 0 to 4 (see

section

5.8.11.1

)

11 QTWIN_XSTAT2

Raw status (XtlStat2_class) Value 0 to 2 (see

section

5.8.11.2 on page 5-36

)

12 QTWIN_XACHNL

Channel that Index sensor is
assigned to

0 - Channel 1
1 - Channel 2

13 QTWIN_SHTSRC

Source shutters

1 - True
Source shutter is open
0 - False
Source shutter is closed

Channels not being used will
return 0.

This is the logical state of the
process, not the actual I/O
relay state