1 raw status, 2 raw status 2, Section – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 116

Advertising
background image

5 - 36

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

5.8.11.1 Raw Status

XtlStat_FLTRINIT . . . . . . . . . . . . . . . 0 ??? CRYSTAL

Frequency filter is filling

XtlStat_OK . . . . . . . . . . . . . . . . . . . . 1 OK CRYSTAL IS GOOD

Averaging filter is full, and rate is meaningful

XtlStat_LOWLIFE . . . . . . . . . . . . . . . 2 LIFE CRYSTAL

Crystal frequency is in range, but not within
bounds

XtlStat_DEAD. . . . . . . . . . . . . . . . . . 3 BAD CRYSTAL

Crystal frequency is out of range

XtlStat_OFF . . . . . . . . . . . . . . . . . . . 4 OFF CRYSTAL

Crystal is not being used

5.8.11.2 Raw Status 2

XS2_normal . . . . . . . . . . . . . . . . . . . 0

Crystal is being used normally

XS2_hold . . . . . . . . . . . . . . . . . . . . . 1

Crystal is in "thick hold" mode, because
substrate shutter is closed or the crystal is in
use as a backup crystal

XS2_coast . . . . . . . . . . . . . . . . . . . . 2

Crystal is in coasting mode; rate and
thickness values are not live (measurement
based) but are incremented/simulated during
time power mode