5.7 Windows/SMDP Server Implementation

SMDP server on Windows message timeout is typically 150 milliseconds. That is,
after the host sends a command, it waits 150 milliseconds for a reply. If no complete
reply is returned from STM-2XM in that time, the message is considered lost. This
could indicate the following:

STM-2XM is offline

The cable is disconnected

The baud rate is invalid/mismatched

There is noise corrupting the packet/checksum integrity

After 150 milliseconds have elapsed, the host can try again.

NOTE: Once a valid connection to STM-2XM is made, the command/response

exchange will proceed at the fastest speed, sending a new command as
soon as the prior answer has come back to the host.

If there is a loss of a packet due to noise, software error, or power loss to STM-2XM,
the loss will be detected when a valid reply packet is not returned within the 150
millisecond timeout.

Retries or resending commands can continue to take place until a valid reply is
returned once the 150 millisecond timeout has been reached.