5 - 23

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

33 CSID_SYSPLOT

Plot type, plot_class

0 - plot_none
No plot, plot disabled
1 - plot_rdev
Rate Deviation, Measured -
desired
2 - plot_rate
Rate, measured rate in plot
units
3 - plot_thick
Thickness, autoscaled

34 CSID_SYSPLOTHSCL

Plot horizontal scale time
type h:mm:ss

Value 0:00:00 to 9:59:59, used
in intervals rounded up to
nearest 5 seconds modulo

For example, 0:0:0 to 0:0:5
acts like 5 seconds.
0:00:06 to 0:00:10 acts like 10
seconds.

35 CSID_SYSLCDCT LCD

contrast

Value 10 to 85 sets contrast
voltage, defaults to 70

36 CSID_SYSLCDBT

LCD brightness

Value 1 to 99 sets backlight
brightness in % of full, default
50

37 CSID_SYSBEEP

Beeper ON/OFF

0 - False/BeepOff
1 - True/BeepOn

Defaults off

38 CSID_SYSPROT

Protocol
sysprot_class

0 - sysprot_SMDPH
SMDP PROTOCOL @
115,200 BAUD
1 - sysprot_SMDPM
SMDP PROTOCOL @ 38,400
BAUD
2 - sysprot_SMDPL
SMDP PROTOCOL @ 9600
BAUD
3 - sysprot_sycon
SYCON PROTOCOL @ 9600
BAUD

39 CSID_SYSSMDPADR

SMDP Address

Value 16 to 254

16 defines RS-232
point-to-point,
17 to 254 give RS-485
multidrop network physical
interface at addresses 17 to
254.
NOTE:

Not used in Sycon
protocol.

Table 5-3 E/F command (continued)

#

Command Structure

Comments

Values