2 programming, 1 main menu – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 43

Advertising
background image

3 - 7

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

3.2 Programming

3.2.1 Main Menu

To program STM-2XM, press PROG/ENTER. The MAIN MENU will display.

Figure 3-8 Main menu

Use the control knob to select a menu item, then press PROG/ENTER. Press
CANCEL/BACK to return to the Runtime screen.

Set Active Film . . . . . . . . . . . . . . . . 1 to 15. Sets the active film.

Edit Active Film . . . . . . . . . . . . . . . Lists programmable parameters for the

currently active film. For a description of the
settings on the film page, see

section 3.2.1.1

.

System Settings . . . . . . . . . . . . . . . Displays the SYSTEM PARAMS menu. For a

description of the system settings, see

section 3.2.1.2 on page 3-10

.

Status/Diagnostics . . . . . . . . . . . . . Displays STM-2XM status and diagnostics.

For example, displays the raw frequency of
the crystal, the status of the inputs and
outputs, and other useful information.

Edit Any Film . . . . . . . . . . . . . . . . . Displays a selection box that allows the

choice of the film to edit, and then displays
the settings menu for the selected film.

Setup . . . . . . . . . . . . . . . . . . . . . . . . Displays the HARDWARE SETUP menu,

where STM-2XM system hardware is
programmed, including sensors, shutters,
tooling, and relays. The settings on this menu
must be accurate to allow STM-2XM to
operate properly. For example, if STM-2XM
is programmed for a substrate shutter, and
one does not exist in the system, STM-2XM
will provide false readings because it will not
count any material accumulated on the
substrate during the Ready phase. For a
description of hardware settings, see

section

3.2.1.3 on page 3-12

.