1 setup parameters pane – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 71

Advertising
background image

4 - 11

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

4.3.3.1 Setup parameters Pane

Sensor setup. . . . . . . . . . . . . . . . . . Main parameter that affects the

choices/modes available (see

Table 4-1

).

Substrate shutter setup . . . . . . . . . The parameter that describes the shutter

assembly of the substrate(s) (see

Table 4-2

).

Tooling and Weights

Tooling X1:S1 . . . . . . . . . . . . . . . . . The ratio between substrate thickness using

Source 1 and Sensor 1 thickness (see

section 7.3, Determining Tooling, on page
7-2

).

Tooling X2: S1 . . . . . . . . . . . . . . . . The ratio between substrate thickness using

Source 1 and Sensor 2 thickness (see

section 7.3, Determining Tooling, on page
7-2

).

NOTE: For independent chambers, set this

value to zero.

Table 4-1 Sensor setup parameter

Sensor Setup

Available modes

Rate sampling

capable

Backup

capable

1:Single

unshuttered

Simple

No

No

2:Single shuttered

Simple

Yes

No

3:Dual w/single

shutter

Simple

Yes

Yes

4:Dual unshuttered

Simple, Averaging,

Alloy, Independent

No

If mode is

Simple

5:Dual w/dual

shutters

Simple, Averaging,

Alloy, Independent

Yes

If mode is

Simple

Table 4-2 Substrate shutter setup parameters

Substrate shutter

setup

Shutter Delay

Thick hold

None N

CH1: Substrate

Only for film on Channel 1

CH2: Substrate

Only for film on Channel 2

Shared substrate

Only allows shutter delay for
one film at a time

Dual substrate

Allows dual shutter delay
(alloy shutter delay)