4 specifications, 1 measurement, 2 film parameters – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 20

Advertising
background image

1 - 6

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

1.4 Specifications

1.4.1 Measurement

Sensor Inputs . . . . . . . . . . . . . . . . . . 2

Compatible Sensors . . . . . . . . . . . . . Single/shuttered single or dual QCM sensor

Measurement Frequency Range . . . 6.0 to 5.0 MHz (fixed)

Frequency Resolution . . . . . . . . . . . ± 0.03 Hz @ 6 MHz

Reference Frequency Stability . . . . . ± 2 ppm

Thickness and Rate
Resolution/Measurement . . . . . . . . . ± 0.037 Å @ tooling/density = 100/1

Fundamental frequency = 6 MHz

Measurement Interval . . . . . . . . . . . 0.10 s

1.4.2 Film Parameters

Stored Films . . . . . . . . . . . . . . . . . . . 15

Name . . . . . . . . . . . . . . . . . . . . . . . . 8 characters, 0 to 9, A to Z, _ and [space]

Density . . . . . . . . . . . . . . . . . . . . . . . 0.4 to 99.99 g/cm

3

Z-Ratio . . . . . . . . . . . . . . . . . . . . . . . 0.100 to 9.999

Tooling . . . . . . . . . . . . . . . . . . . . . . . 10 to 999.9%

Setpoint Thickness. . . . . . . . . . . . . . 0.000 to 999.999 kÅ

End Thickness . . . . . . . . . . . . . . . . . 0.000 to 999.999 kÅ

Use Backup Crystal . . . . . . . . . . . . . Yes/No

Setpoint Time . . . . . . . . . . . . . . . . . . 0:00:00 to 9:59:59 h:mm:ss

Deposit Rate . . . . . . . . . . . . . . . . . . 0.0 to 999.9 Å/s

Coast Enable . . . . . . . . . . . . . . . . . . Yes/No

Crystal Life Minimum . . . . . . . . . . . . 0 to 99%

Force Crystal Fail . . . . . . . . . . . . . . . Yes/No

Temporal Average . . . . . . . . . . . . . . 1 to 50

Good Rate Tolerance . . . . . . . . . . . . 0 to 50

Advertising