INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 74

Advertising
background image

4 - 14

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

Digital (relay) outputs

Relay 1 to 8 function . . . . . . . . . . . Determines the function of each of the digital

(relay) outputs (refer to

section 2.2.2 on page

2-4

for details on wiring digital outputs). (See

Table 4-4

.)

Table 4-4 Relay functions

Setting

Decription

0: Forced OFF

Relay is inactive

1: Forced ON

Relay is always active

2: Computer controlled

Relay is controlled with remote communications

3: Front panel buttons, top

Relay is toggled using Sensor 1 OPEN/CLOSE buttons

4: Front panel buttons, bottom

Relay is toggled using Sensor 2 OPEN/CLOSE buttons

5: Channel 1 source shutter

Relay activates Channel 1 source shutter

6: Channel 2 source shutter

Relay activates Channel 2 source shutter

7: Channel 1 substrate shutter

Relay activates Channel 1 substrate shutter

8: Channel 2 substrate shutter

Relay activates Channel 2 substrate shutter

9: Sensor 1 (X1) active

Relay activates Sensor 1 shutter

10: Sensor 2 (X2) active

Relay activates Sensor 2 shutter

11: Sensor 1 (X1) failed

Relay activates when Sensor 1 fails

12: Sensor 2 (X2) failed

Relay activates when Sensor 2 fails

13: Channel 1 setpoint time

Relay activates when Channel 1 setpoint time has elapsed

14: Channel 2 setpoint time

Relay activates when Channel 2 setpoint time has elapsed

15: Channel 1 setpoint thick

Relay activates when Channel 1 setpoint thickness has been
reached

16: Channel 2 setpoint thick

Relay activates when Channel 2 setpoint thickness has been
reached

17: Channel 1 good rate

Relay activates when Channel 1 rate is steady and within tolerance

18: Channel 2 good rate

Relay activates when Channel 2 rate is steady and within tolerance

19: Complement of #9

Relay is Normally Closed and opens to use Sensor 1

20: Complement of #10

Relay is Normally Closed and opens to use Sensor 2

Advertising