Metrohm tiamo 2.4 Manual User Manual

Page 1126

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1110

■■■■■■■■

tiamo 2.4

NOTE

Only instruments for a one-channel instrument setup can be selected.
For a wavelength calibration of a two-channel instrument setup, both
instruments have to be reconfigured in such a way that they have nei-
ther an instrument for the reference channel nor an instrument for the
measuring channel (Instrument for reference channel = not defined
then has to apply to the instrument for the measuring channel). In addi-
tion, the light source should be set to continuous, and the measuring
setup should be optimized for an AvaLight-CAL light source.

Appearance
The command has the following appearance:

Parameters
The parameters for the CAL Spec command are set on the following
three tabs:

■

General/Hardware
Parameters for instrument.

■

Measuring parameters
Parameters for setting the measurement procedure.

■

Calibration wavelengths
Entry of the reference wavelengths, definition of the behavior of the
calibration.

Command variables
The following command variables are generated by the CAL Spec com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable designation':

Identification

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.C0

Coefficient c0 of the calibration curve

.C1

Coefficient c1 of the calibration curve

.C2

Coefficient c2 of the calibration curve