5 Method

tiamo 2.4

■■■■■■■■

857

NOTE

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

If this command is already started (e.g. via Automatic start after
sample addition
), the SEND command will not have any effect.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.

5.6.3.7.8

BRC - Additional evaluations

Tab: Method

BRC Properties... Additional evaluations

Command name

Name of the command.

Entry

25 characters

The following additional method for evaluation of the titration curves can
be activated and defined on this tab:

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fixed measured quantity (mea-
sured value
, time or bromine) for the fixed endpoint from the measur-
ing point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.7.10, page 859).