5 Method

tiamo 2.4

■■■■■■■■

891

Appearance
The command has the following appearance:

Parameters
The parameters for the STAT U command are set on the following eight
tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Control parameters
Parameters for the control point.

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definition of conditions which cause the titration to stop.

■

Monitoring
Definition of the monitoring of measured value, dosing rate and tem-
perature.

■

Evaluations
Definition of methods for evaluation of the titration curves.

■

Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated by the STAT U com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

.BLV

Blank value of the sensor used for the command (only
for ISE sensors)

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started