Metrohm tiamo 2.4 Manual User Manual

Page 1355

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4

■■■■■■■■

1339

Creating new template or editing template
With <New> or <Edit> the dialog window New template or Edit tem-
plate
opens for entering a new template or for editing an existing tem-
plate.

Signal name

Name of the pattern for the output signal.

Entry

25 characters

Output signal

Entry of the bit pattern for the output signal with exactly 14 characters.

It is possible to enter the characters

0 = Line inactive,

1 = Line active,

* = Any line status and

p = Set pulse

Selection

Bit pattern with exactly 14 characters (0, 1, *,
p)
| **************

Default value

**************

The output lines are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.

NOTE

We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line conditions.

Pulse length

Duration of the transmitted pulse.

Input range

100 - 1000 ms

Default value

200 ms

Advertising