4 kft upol - control parameters, Control parameters – Metrohm tiamo 2.4 Manual User Manual

Page 831

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4

■■■■■■■■

815

Start volume

Start volume

Volume to be dosed with the indicated dosing rate before the start of the
titration.

Titrando
Input range

0.00000 - 9,999.99 mL

Default value

0.00000 mL

Titrino
Input range

0.00 - 999.99 mL

Default value

0.00 mL

Dosing rate

Rate at which the start volume is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Input range

0 - 999999 s

Default value

0 s

5.6.3.5.3.4

KFT Upol - Control parameters

Tab: Method

KFT Upol Properties... Control parameters

Command name

Name of the command.

Entry

25 characters