5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

740

■■■■■■■■

tiamo 2.4

Parameters
The parameters for the SET Ipol command are set on the following eight
tabs:

■

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

■

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

■

Control parameters
Control parameters for endpoint 1 and endpoint 2 (measured values
for the endpoints, titration rate, user-defined parameters, stop crite-
rion).

■

Titration parameters
Parameters for the run of the titration.

■

Stop conditions
Definition of conditions which cause the titration to stop.

■

Conditioning
Enable/disable conditioning. Definition of start drift, drift correction
and stop conditions.

■

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

■

Additional measured values
Definition of additional measured values of other measuring com-
mands, which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated by the SET Ipol com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated