3 test login for password monitoring by windows, 4 audit trail / modifications, Audit trail/changes – Metrohm tiamo 2.4 Manual User Manual

Page 1334: Audit trail / changes

Advertising
background image

6.2 Administration

■■■■■■■■■■■■■■■■■■■■■■

1318

■■■■■■■■

tiamo 2.4

Input range

1 - 999 days

Default value

365 days

6.2.2.3

Test login for password monitoring by Windows

Tab: Configuration

Tools Security settings Login/Password protec-

tion

Test login

If the Password monitoring by Windows option (see Chapter 6.2.2.2,
page 1313)
is activated, then the Windows password of the user must be
entered in this dialog window. The test login is used to check whether the
user name matches the Windows user name.

User

Shows the current user name.

Password

Entry of the Windows password.

It is only when the test login is successful that the Password monitoring
by Windows
can be used.

6.2.2.4

Audit Trail / Modifications

Tab: Configuration

Tools Security settings Audit Trail/Modifications

The recording of the Audit Trail is switched on and off on the Audit
Trail/Modifications
tab. When a method, a determination or sample
data is modified, it can also be defined whether or not a reason and a
comment must be given for the modification.

Audit Trail

Enable Audit Trail

If this check box is activated then all program actions will be automati-
cally recorded; these are defined as Audit Trail actions.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Modifications

Comment on modification of methods

If this check box is activated, then each time that a method is modified a
modification reason and a modification comment must be entered; these
are saved in the method and shown in the method history. The reason
and comments are also recorded in the Audit Trail.