11 met u - fixed endpoint evaluation, 4 met ipol, 1 met ipol - overview – Metrohm tiamo 2.4 Manual User Manual

Page 636

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

620

■■■■■■■■

tiamo 2.4

5.6.3.3.3.11

MET U - Fixed endpoint evaluation

Dialog window: Method

▶ MET U ▶ Properties... ▶ Additional evaluations ▶ Fixed

endpoint evaluation

▶ [New]/[Properties] ▶ Fixed endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-2000.0 - 2000.0 mV

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.3.4

MET Ipol

5.6.3.3.4.1

MET Ipol - Overview

Dialog window: Method

▶ MET Ipol ▶ Properties... ▶ MET Ipol - 'Command name'

Command for Monotonic equivalence point titrations with voltamet-
ric measurement (selectable polarization current).

Instruments
This command can be executed with the following devices.

Titrando: 808, 809, 835, 836, 852, 857, 888, 904, 905, 906, 907

Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance: