4 determination series, 1 determination series - overview, 2 determination series - controls – Metrohm tiamo 2.4 Manual User Manual

Page 162: Determination series, Start/stop series, Interrupt/resume determination, Interrupt/resume series

Advertising
background image

3.6 Run subwindow

■■■■■■■■■■■■■■■■■■■■■■

146

■■■■■■■■

tiamo 2.4

If the option Comment on modification of sample data (live) is
enabled in the security settings, then this window must be opened before
saving. A reason and a comment on the modification must be entered
here (see Chapter 6.2.2.4, page 1318).

Reason

Selection from the standard reasons for the category Modifications of
sample data
(see Chapter 6.2.2.6, page 1320) that are defined in the
security settings.

Selection

Selection from default reasons

Comment

Entry of remarks about the modifications to the sample data.

Entry

1000 characters

3.6.4

Determination series

3.6.4.1

Determination series - Overview

Tab: Workplace

Run Determination series

Parameters and sample data for determination series can be edited and
run functions can be triggered on the Determination series tab in the
Run subwindow. It contains the following elements and functions:

Controls
Buttons for starting, stopping, interrupting and resuming determina-
tions and series.

Status display
Shows the current status of the workplace.

Sample data
Entry of sample data for the next determinations.

Properties
Definition of the properties for the Determination series tab.

3.6.4.2

Determination series - Controls

Tab: Workplace

Run Determination series

The following controls are located on the Determination series tab:

Start series (shortcut: Ctrl G)

Start of the first determination of a sample series. The button is present in
the status READY (no determination started). If the start can be triggered,
then it will be green, otherwise light gray. For methods that have condi-
tioning, conditioning will start first, followed automatically by the determi-
nation (status COND READY).

Advertising