7 Devices

tiamo 2.4

■■■■■■■■

1579

Measuring input 1/2 (iConnect)

ADC type

Shows the type of analog-digital converter.

Serial number

Shows the serial number of the measuring input interface.

Temperature sensor

Selects the type of temperature sensor connected to the measuring input.

With the 856 Conductivity Module displayed only.
Selection

Pt 1000 | NTC

Default value

Pt 1000

R (25 °C)

Nominal resistance of connected NTC sensor.

Only visible with temperature sensors of the NTC type.
Input range

1000 - 99999 Ohm

Default value

30000 Ohm

B value

Material constant of the NTC resistance referred to measuring the resist-
ance at 25 °C and 50 °C.

Only visible for temperature sensors of the NTC type.
Input range

1000 - 9999

Default value

4100

7.17.4

Properties - MSB #

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

MSB #

On the tabs MSB # the properties of the connector and the devices con-
nected to are displayed.

Request for dosing device preparation
Selection when the request for carrying out the command PREP (see
Chapter 5.6.6.5.1, page 1207)
(prepare) is to be shown for the dosing
device connected to the MSB.