5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

906

■■■■■■■■

tiamo 2.4

since

Specifies when the stop time begins to count (will not be displayed if the
Stop time is disabled):

Selection

Start | Control point first reached | Last dosing

Default value

Start

Start
The stop time starts after the start volume has been dosed and after
the following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i.e., the
stop time is reset to zero again after each dosing step.

Stop rate

Rate at which the dosing cylinder is to be filled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing unit used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

off

Default value

off

Titrino
Input range

0.01 - 150.00 mL/min

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum