2 properties - colorimetric sensors, 3 properties - calibration data – Metrohm tiamo 2.4 Manual User Manual

Page 1443

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4

■■■■■■■■

1427

Limit values
Definition of the limit value monitoring for outliers and the coefficient
of determination.

History
Display of the last 10 calibration values.

6.8.3.2

Properties - Colorimetric sensors

Tab: Configuration

Colorimetric sensors Edit Properties... Colorimetric

sensor

Name

Name of the sensor.

Entry

24 characters

Comment

Possibility to enter remarks about the sensor.

Entry

24 characters

Default value

'empty'

6.8.3.3

Properties - Calibration data

Tab: Configuration

Colorimetric sensors Edit Properties... Calibration

data

Wavelength

Wavelength at which the calibration of the instrument has been carried
out.

Input range

100.0 - 2000.0 nm

Default value

400.0 nm

c0

Calibration coefficient of zero order.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0

c1

Calibration coefficient first-order.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0

c2

Calibration coefficient second-order.

Input range

-1E+99 - 1E+99 (max. 10 significant places)

Default value

0