5 Method
tiamo 2.4
■■■■■■■■
615
Lower limit
Value for the lower limit of the window.
Window = Measured value
Input range
-2000.0 - 2000.0 mV
Default value
-2000.0 mV
Window = Volume
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Window = Time
Input range
0 - 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value
Input range
-2000.0 - 2000.0 mV
Default value
2000.0 mV
Window = Volume
Input range
0.00000 - 9999.99 mL
Default value
9999.99 mL
Window = Time
Input range
0 - 999999 s
Default value
999999 s
Gran evaluation
NOTE
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accor-
dance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 932).
Initial volume
Volume which must be present in the measuring vessel before the com-
mand is started.