5 Method

tiamo 2.4

■■■■■■■■

615

Lower limit

Value for the lower limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

-2000.0 mV

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Window = Time
Input range

0 - 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value
Input range

-2000.0 - 2000.0 mV

Default value

2000.0 mV

Window = Volume
Input range

0.00000 - 9999.99 mL

Default value

9999.99 mL

Window = Time
Input range

0 - 999999 s

Default value

999999 s

Gran evaluation

NOTE

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accor-
dance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 932).

Initial volume

Volume which must be present in the measuring vessel before the com-
mand is started.