5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1174

■■■■■■■■

tiamo 2.4

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | Titrando

Default value

Titrando

Dosing device

Dosing device

Selection of the dosing device to be used for the dosing. All the dosing
device connections which are possible with the selected device type are
displayed.

Titrando

Selection

1 | 2 | 3 | 4

Default value

1

855

Selection

1 | 2 | 3

Default value

1

Titrino
Selection

internal D0

736, 751, 799
Selection

internal D0 | external D1 | external D2

Default value

internal D0

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Entry

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.