Metrohm 732 IC Detector User Manual

Page 105

Advertising
background image

4 Operation

732 IC Detector / 733 IC Separation Center

98

>PARAM/detector
full scale: 1.00 mS/cm

0.05

µ

S/cm

10 mS/cm

Full-scale range (operating range)
The full-scale range (operating range) sets the desired

sensitivity for the display and analog output of the

measuring signal during the recording of a chroma-

togram. The following points should be noted:

The possible values of the full-scale range which can

be selected with the <SELECT> key depend on the

preset measuring range

"

range

"

. This value can be

reduced in a total of 11 stages by a factor 1 (lowest

sensitivity) up to maximum a factor of 2000 (highest

sensitivity).

The full-scale range is primarily used to increase the

sensitivity for the recording of the chromatogram

following the electronic background compensation of

the eluent sensitivity (auto-zero function, see also

description in section 4.6.3).

For the selected full-scale range (FS), the linearity of

the conductivity measurement is assured in the

range -1.5 FS

+1.5 FS. If the measured value vio-

lates these limits, overload display 88 lights up, but

the measured value is still displayed. If the measured

value lies outside

±

1.8 FS, overload display 88 starts

to flash, at the same time display 11 shows the abso-

lute conductivity in place of the auto-zero value.

The full-scale range determines the limits for the

output of the analog signal at analog outputs 11

11

(0

1 V) and 12

12 (0

10 mV). The polarity of the out-

put signal can be changed at any time with the pa-

rameter

"

polarity

"

to

"

+

"

(0

+FS) or

"

"

(0

–FS).

0

R

1.1 R

0

+ FS

– FS

+ 1.5 FS

+ 1.8 FS

– 1.5 FS
– 1.8 FS

0

+

0 V

1 V / 10 mV

1 V / 10 mV

Measuring range

Full-scale range FS

Analog output

Conductivity

Display

"

--------

"

Overload lit up

Overload flashes

Overload lit up

Overload flashes