4.6.4

<ZERO OFF> key

ZERO OFF

4

The <ZERO OFF> key is used to switch off the auto-zero function. The

auto-zero function can be triggered using the <ZERO OFF> key in the

standby mode or when a program is running, but not in the edit mode.
When the green LED 66 in the <ZERO> key goes out, this shows that

the electronic background compensation is no longer active.

4.6.5

<MARK> key

MARK

0

The <MARK> key is used to trigger a marking signal. In addition to the

measuring signal, a signal of ca. 10% of the full-scale range is output-

ted briefly at the analog output. This allows, e.g. the start of a chroma-

togram to be marked during recording on a recorder.
Marking can be triggered with the <MARK> key in the standby mode

or when a program is running, but not in the edit mode.

4.6.6

<PUMP R/S> key

PUMP R/S

8

The <PUMP R/S key is used for on/off switching of the pump drive of

the 709 IC Pump. A requirement for this is that the 709 IC Pump is

connected to the 732 IC Detector and the remote control is switched on

at the pump with the <EXT.> key (see section 2.6.1).
With the <PUMP R/S> key the pump drive of the 709 IC Pump can be

switched on or off in the standby mode or when a program is running,

but not in the edit mode. When the pump drive is switched on, the red

LED above the <R/S> key on the pump lights up.
The settings of the pump parameters on the 732 IC Detector 732 are

described under the <CONFIG> key in section 4.4.2, the pump status

messages in the standby mode in section 4.3.2.