2.7 Precolumns

732 IC Detector / 733 IC Separation Center

31

2.7.4

IC anion precolumn SUPERSEP

The 6.1009.010 IC Anion Precolumn SUPERSEP has two connections

for steel or PEEK capillaries and is installed as follows:

1 Connect precolumn

•

Remove end caps from the precolumn.

•

Fit connectors to column connection capillary 67

67 mounted on

the injection valve (see section 2.5).

•

Screw precolumn to column connection capillary 67

67.

•

Cut a piece from the 6.2620.020 Steel capillary or the

6.1831.010 PEEK capillary as short as possible and fit with

connectors (see section 2.5).

•

Fasten the prepared capillary to the other end of the precol-

umn.

2 Rinse precolumn

•

Place a beaker beneath the outlet capillary of the precolumn.

•

Switch on 709 IC Pump and rinse precolumn with eluent for

ca. 10 min.

•

Switch off IC pump.