Metrohm 732 IC Detector User Manual
Page 161

5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
154
5.4.7
Check remote interfaces
This diagnostic step can be used to check all outputs and inputs of the
two remote interfaces 17
17 "Remote" and 14
14 "733 IC Separation Center"
for functionality.
•
Prepare instrument for diagnostic test (see section 5.4.2).
•
If need be, press <9> key repeatedly until
diagnosis
>diag/IO test
•
<ENTER>
>diag/IO test
>diag/IO test/remote
•
<ENTER>
>diag/IO test/remote
IO connector?
•
Without switching off the instrument, plug the 3.496.8510 Test plug
into socket "Remote".
•
<ENTER>
The test runs automatically. If no fault is found, the following appears:
>diag/IO test/remote
IO test ok ..
•
Remove test plug and <ENTER>.
>diag/IO test
>diag/IO test/733 Sep.C.
•
<ENTER>
>diag/IO test/733 Sep.C.
IO connector?
•
Without switching off the instrument, plug 3.496.8510 Test plug into
socket "733 IC Separation Center".
•
<ENTER>
The test runs automatically. If no fault is found, the following appears:
>diag/IO test/733 Sep.C.
IO test ok ..
•
Remove test plug and <ENTER>.
diagnosis
>diag/internal test