5.4 Diagnostic tests
732 IC Detector / 733 IC Separation Center
153
5.4.6
Check RS232 interfaces
With this diagnostic step all outputs and inputs of the two RS232 inter-
faces 16
16 "RS 232" and 15
15 "709 IC Pump" can be checked for function-
ality.
•
Prepare instrument for diagnostic test (see section 5.4.2).
•
If need be, press <9> key repeatedly until
diagnosis
>diag/RS test
•
<ENTER>
>diag/RS test
>diag/RS test/RS232
•
<ENTER>
>diag/RS test/RS232
RS connector?
•
Without switching off the instrument, plug the 3.496.8480 Test plug
into socket "RS232".
•
<ENTER>
The test runs automatically. If no fault is found, the following appears:
>diag/RS test/RS232
RS test ok
•
Remove test plug and <ENTER>.
>diag/RS test
>diag/RS test/709 Pump
•
<ENTER>
>diag/RS test/709 Pump
RS connector?
•
Without switching off the instrument, plug the 3.496.8480 Test plug
into socket "709 IC Pump".
•
<ENTER>
The test runs automatically. If no fault is found, the following appears:
>diag/RS test/709 Pump
RS test ok
•
Remove test plug and <ENTER>.
diagnosis
>diag/IO test