4.3 Displays in the standby mode

732 IC Detector / 733 IC Separation Center

83

stopped: YY-MM-DD HH:MM

The 709 IC Pump was stopped at the time

indicated owing to violation of the shutoff

limit

"

Pmin

"

or

"

Pmax

"

. To restart a pump

stopped in this manner, the pump keys

<EXT.> and <R/S> must be pressed in

succession and then <EXT.> pressed

again (see also section 4.5.1).

pump not responding

This message appears when

•

the 709 IC Pump is not connected to the

732 IC Detector,

•

the 709 IC Pump is not switched on,

•

the remote control of the 709 IC Pump is

not switched on.

+0.004

µµS/cm

13.5 min

prog.type cycle no. XXX

Status messages for program type

These status messages appear only with
"

PROGRAM/parameters/status = active

"

.

prog.type cycle no. XXX

Display of the program type

"

cycle

"

and the

cycle number (see section 4.7.1).

prog.type inject

Display of the program type

"

inject

"

(see

section 4.7.1).

prog.type remote

Display of the program type "

remote

" (see

section 4.7.1).

+0.004

µµS/cm

13.5 min

next step ready

Status messages for program step

These status messages appear only with
"

PROGRAM/parameters/status = active

"

.

next step ready

The program is ready for a new start (see

section 4.7.1).

next step XXX.X min

Display of the time when the next program

step will be executed.

next step ---

The program has been set to

"

inactive

"

, a

new start is not possible.

next step

ì

The program continues to run indefinitely (no

"

end

"

command available).

Advertising