3 Operating tutorial

732 IC Detector / 733 IC Separation Center

50

151.2

µµS/cm 1.2 min

pump ready

The bottom line of the display shows the status message

for the 709 IC Pump. If the remote operation is switched on

correctly (see point 4), the status message

"

pump ready

"

now appears.
If the IC Pump is operated independently, the message

"

pump not responding

"

appears instead.

SELECT

Press the <SELECT> key again.

151.2

µµS/cm 0.9 min

method DEFAULT

The bottom line of the display shows the status message

for the loaded method once again. The

"

DEFAULT

"

method

in which all parameters are set to the initial values is loaded

as standard. You have now returned to the status mes-

sages displayed at the start.

The next point of the tutorial is concerned with the funda-

mentals of data entry using the settings of the instrument

configuration required for the example.

6 Set configuration

7

CONFIG

Press the <CONFIG> key to open the main menu for the

instrument configuration.

CONFIG
>CONFIG/detector

The name of the selected key appears in the top line of the

display, the title of the submenu

"

detector

"

, which con-

tains various inquiries for the conductivity detector, is

shown in the bottom line.
The main menu of the <CONFIG> key contains several

such submenus, which are selected by repeated pressing

of this key. Each submenu has a title marked by

"

>

"

.

ENTER

Now press the <ENTER> key.

>CONFIG/detector
thermostat: 35 °C

This moves you from the title to the individual inquiries of

the submenu

"

detector

"

; the title continues to be shown in

the top line. The operating temperature of the conductiv-

ity cell appears in the display as the first inquiry.
Please note the character

"

:

"

with this display. It always

appears when values can not be entered using the numeric

keys, but must be selected from preset values using the

<SELECT> key. However, for our example the preset value

of 35 °C does not have to be changed.

ENTER

Confirm the set operating temperature by pressing the

<ENTER> key. The displayed value will be accepted and a

switch made to the next inquiry.