Metrohm 732 IC Detector User Manual

Page 56

Advertising
background image

3.3 Putting into operation

732 IC Detector / 733 IC Separation Center

49

4 Switch on 732 IC Detector

151.2

µµS/cm 0.0 min

method DEFAULT

Switch on 732 IC Detector with mains switch 99 at the rear of

the instrument.

After the instrument has been switched on, display 11 lights

up and shows the status messages for the basic instrument

condition.
The 732 IC Detector is now in the standby mode of con-

ductivity measurement. The displayed values have the fol-

lowing meaning:

Instead of the loaded method, other status messages can

be selected which are then continuously displayed in the

standby mode in the bottom line of display 11. Proceed as

follows:

5 Select status message

SELECT

Press the <SELECT> key.

151.2

µµS/cm 0.4 min

1995-09-11 14:15:27

The bottom line of the display shows the current date and

the current time. Please refer to point 6 to see how to

change the date and time.

SELECT

Press the <SELECT> key again.

151.2

µµS/cm 0.6 min

full scale 1.00 mS/cm

The bottom line of the display now shows the full-scale

range.

SELECT

Press the <SELECT> key again.

151.2

µµS/cm 0.9 min

abs.cond. 151.2

µµS/cm

The bottom line of the display now shows the current

measurement of the absolute conductivity, which is

identical to the value already shown in the top line on the

left.

SELECT

Press the <SELECT> key again.

151.2

µµS/cm 0.2 min

full scale 1.00 mS/cm

Current time

after switching

on instrument

Measured

absolute

conductivity

Display of the

loaded method