Metrohm 732 IC Detector User Manual

Page 131

Advertising
background image

4 Operation

732 IC Detector / 733 IC Separation Center

124

4.8.3

<REPORT> key

REPORT

2

The <REPORT> key is used for the output of reports on an external

printer or PC connected to the RS232 interface.
The <REPORT> key opens the following inquiry:

XXXXXX

µµS/cm XXXX min

report: all

all, config, event,

method, method list,

param, program, pump

Selection of the report

Selection of the report which should be outputted:

all

All reports in the order "

config,

event,

method,

method list,

pump

"

config

Configuration report (parameters of the

<CONFIG> key)

event

Event report (parameters of the

<EVENT> key)

method

Method report (comprises method name,

parameter and program report)

method

Method list (details of stored

list

methods)

param

Parameter report (parameters of the

<PARAM> key)

program

Program report (parameters of the

<PROGRAM> key)

pump

Report of the pump parameters

Example of "config" report

Configuration for 732 IC Detector with attached Seiko printer and 733.0010

IC Separation Center with 1 injection valve

CONFIG
>CONFIG/detector
thermostat: 35 °C
'zero' unit:

µ

S/cm

cell constant 16.7 /cm
>CONFIG/printer
id.1 PRP-X100
id.2 drinking water
print header: once
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: time
time interval 300.0 s
stop time 60 min
date&time: off
>CONFIG/auxiliaries
run number 0
number of cycles 1
>CONFIG/aux/event
enabled in program: off
date 1995-10-10
time 11:38:20
dialog: english
device label
program 732.0012

>CONFIG/aux/beeper
status: on
repeat time 60 s
>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
control via RS: on
>CONFIG/RS settings 709
baud rate: 9600
>CONFIG/733 IC Sep.Cent.
valve A
control: no restriction
>CONFIG/733/valve A
status inject
trigger: inject
--------------