Metrohm 732 IC Detector User Manual
Page 27

2 Installation
732 IC Detector / 733 IC Separation Center
20
49
49
Ferrule (6.2620.010)
50
50
Pressure screw (6.2620.000)
51
51
Capillary
6.2620.020 Steel capillary or
6.1831.010 PEEK capillary
52
52
Compression fitting
(6.2744.010)
2.5.3
PEEK connectors
For the connection of 6.1822.010 PEEK capillaries (i.D. = 0.3 mm) or
6.1822.010 PTFE microcapillaries, the 6.2744.010 PEEK compres-
sion fittings are used. Proceed as follows:
1 Mount compression fitting
Slide a compression fitting 52
52 (6.2744.010) over the end of the
capillary 51
51 to be fastened as shown in Fig. 9.
2 Insert capillary in connection
Push capillary end in the corresponding connection as far as it
will go (to avoid dead volume).
3 Tighten compression fitting
Tighten compression fitting 52
52 by hand (never use tools).
50
50
49
49
51
51
52
52
Fig. 9:
Connectors for capillaries