Metrohm 732 IC Detector User Manual
Page 73

3 Operating tutorial
732 IC Detector / 733 IC Separation Center
66
>METHOD/store
name: Water
The display shows the complete method name
"
Water
"
for
confirmation.
ENTER
⇒
Press the <ENTER> key to confirm the method name. The
parameters are now stored in the instrument under the
name
"
Water
"
.
+0.000
µµS/cm 28.4 min
full scale: 4.00
µµS/cm
The 732 IC Detector is now again in the standby mode and
the display shows the status messages.
A method report of the method "Water" just stored can be
outputted with the <REPORT> key (see section 4.8.3) and
has the following appearance:
732 IC Detector 01104 732.0012
date 1995-09-18 time 11:03:47
METHOD
Method name Water
date: 1995-09-18 10:39:10
PARAM
>PARAM/detector
range: 200
µ
S/cm
full scale: 4.00
µ
S/cm
temp.coeff.: 2.5 %/°C
>PARAM/analog output
polarity: +
offset: 0 %fs
damping: off
>PARAM/plot
autom.start: off
time interval 1.0 s
time scale 10.0 mm/min
time scale label: rel
stop time off min
left: 0.000
µ
S/cm
right: 10
µ
S/cm
>PARAM/709 IC Pump
flow 2.00 mL/min
Pmax 10.0 MPa
Pmin 1.0 MPa
flow corr. 1.00
PROGRAM
--------------