3.3 Putting into operation

732 IC Detector / 733 IC Separation Center

51

>CONFIG/detector
'zero' unit:

µµS/cm

The next inquiry of the submenu

"

detector

"

appears,

namely the unit for display of the auto-zero value.
This parameter also contains the character

"

:

"

. The

<SELECT> key can be used to select the other values

"

%fs

"

(% full scale) or

"

mV

"

. For our example, however, the

preset value

"

µ

S/cm

"

need not be changed.

ENTER

Confirm the set unit by pressing the <ENTER> key.

>CONFIG/detector
cell constant 16.7 /cm

The cell constant of the conductivity cell in the detector

block appears as the next inquiry of the submenu

"

detector

"

. Each detector block has a characteristic cell

constant which is determined in the factory and printed on

the block. To ensure that the absolute conductivity is cor-

rectly displayed, this value must be entered on first time

startup.
Please note that this parameter does not contain the

character

"

:

"

. This means that here values must be en-

tered using the numeric keys. Hence press the corre-

sponding numeric keys in the correct sequence to enter the

value printed on the detector block. During the entry you

can always return to the default value by pressing the

<CLEAR> key and restart the entry.

>CONFIG/detector
cell constant 17.1 /cm

A cell constant of 17.1 /cm is entered here as an example.

ENTER

Confirm the cell constant just entered by pressing the

<ENTER> key.

CONFIG
>CONFIG/printer

As the previous inquiry was the last of the submenu

"

detector

"

, a switch is now made automatically to the title

of the next submenu

"

printer

"

, which contains various in-

quiries concerning the printout on an external printer. As no

printer is attached in our example, the next submenu can

be selected directly.

7

CONFIG

To do this, press the <CONFIG> key.

CONFIG
>CONFIG/print meas.value

The title of the submenu

"

print meas.value

"

appears in

the bottom display line. The submenu contains various in-

quiries for the measured value printout on an external

printer, but again we are not interested in this at the present

time.

7

CONFIG

Press the <CONFIG> key again.