5 check display – Metrohm 732 IC Detector User Manual

Page 159

Advertising
background image

5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

152

Key table

Code

Key

Code

Key

0

<9 / METHOD>

12

<CLEAR>

1

<8 / PUMP R/S>

13

<SELECT>

2

<7 / CONFIG>

14

<3 / PRINT>

3

<ZERO>

15

<2 / REPORT>

4

<PARAM>

16

<1 / PLOT>

5

17

6

<6 / PROGRAM>

18

<ENTER>

7

<5 / EVENT>

19

<QUIT>

8

<4 / ZERO OFF>

20

<-/+ / è>

9

<PROG R/S>

21

<. / ç>

10

<FULL SCALE>

22

<0 / MARK>

11

23

5.4.5

Check display

This diagnostic step is used to check all light emitting diodes and the

display for functionality.

Prepare instrument for diagnostic test (see section 5.4.2).

If need be, press <9> key repeatedly until

diagnosis
>diag/display test

<ENTER>

After the <ENTER> key has been pressed, the program automatically runs through a
test sequence for a visual check on the light emitting diodes and the display.

The light emitting diodes for OVERLOAD, THERMOSTAT, ZERO and
PROG R/S are switched on for a certain time.

⇒ Power on test pattern (each pixel active) appears.

Both lines of the display are cleared.

Both lines of the display are written to with the characters “#”, “H”
and finally with “I”.

Both lines are written to from right to left with the continuous moving
display „0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZ“.

⇒ The light emitting diodes for OVERLOAD, THERMOSTAT, ZERO and

PROG R/S flash in succession for a brief time.

The test sequence can be held by pressing any key (with the excep-

tion of <9>) and restarted.

The test is quit with the <9> key.

diagnosis
>diag/RS test

Advertising