4.7 Programming
732 IC Detector / 733 IC Separation Center
109
test
Program test
The test status is used to test a program. Af-
ter the program status has been set to
"
test
"
and return to the standby mode, a
switch is automatically made to the test
mode, which runs as follows:
•
The bottom line of display shows the
message
"
test step
"
with details of the
time
"
X.XXX min
"
when the first program
step should be executed.
•
Pressing the <ENTER> key executes the
actions associated with the program
step.
•
The display shows the next program
steps in turn which are triggered with
<ENTER>.
•
After execution of the last program step,
the message
"
program end
"
appears
and must be confirmed with <ENTER>.
The instrument then returns to the
standby mode and the program status is
set to
"
disabled
"
.
>PROGRAM
>PROGRAM/edit
Edit program
This submenu is used for the entry of new program steps or
to edit existing program steps and their program points.
>PROGRAM/edit
time: x.x min
0.0
…
999.9 min
Edit program step
Under this inquiry new program steps can be entered or
existing ones selected, changed and deleted. Each
program step includes a definite time (two program
steps with the same time can not be entered). A pro-
gram can contain up to 20 program steps.
Enter new program step
A new program step can be entered when
"
x.x min
"
for the time appears in the display. The new time en-
tered with the numeric keys must be confirmed with the
<ENTER> key, the individual program points which
should be executed at this time can then be entered.
Select existing program steps
Program steps already entered can be selected with the
<SELECT> key. After the <ENTER> key has been
pressed, the individual program points of the selected