4.2 Fundamentals of the operation

732 IC Detector / 733 IC Separation Center

73

Main function keys of 732 IC Detector

Key

Standby mode

Edit mode

ZERO

Trigger auto-zero

Automatic zeroing of the current

conductivity (the green LED in the

key lights up).

Trigger auto-zero

Automatic zeroing of the current

conductivity (the green LED in the

key lights up).

PARAM

Open parameter menu

Opens the main menu for setting the

parameters for the conductivity de-

tector, analog output, graphics plot

and 709 IC Pump.

All settings which are set in the

parameter menu can be stored to-

gether with the program as a

method.

Select submenu or parameters

Selection of the next submenu in the

main menu.

Selection of the next parameter in the

submenu.

SCALE

FULL

Set full-scale range

Direct selection of the full-scale range

(value flashes).

Selection of the value with the <ç>

and <è> keys.

Return to standby mode

Acceptance of the displayed value for

the full-scale range and return to the

standby mode.

R/S

PROG

Start/stop program

LED in key dark:

Program inactive; No function

(program start not possible until pro-

gram status =

"

active

"

).

LED in key lit up:

Program active; Program start with

program type

"

cycle

"

.

LED in key flashes:

Program running; Program stop

with all program types.

No function