2 remote interfaces, Remote" interface, 733 ic separation center" interface – Metrohm 732 IC Detector User Manual
Page 190: 1 "remote" interface

6.2 Remote interfaces
732 IC Detector / 733 IC Separation Center
183
6.2 Remote interfaces
The 732 IC Detector has the two remote interfaces 14
14 "733 IC Separa-
tion Center" and 17
17 "Remote" (see Fig. 3) which each have 8 input and
8 output lines.
6.2.1 "Remote" interface
The remote interface 17
17 "Remote" is used for the connection of any
type of external device (see section 2.9.8) and has the following pin as-
signment:
732 IC Detector
Pin
Function
Inputs
+5V
21
Triggering the print command
Equivalent to pressing the <PRINT> key (see section 4.8.1).
H
L
t
P
> 100 ms
+5V
+5V
+5V
+5V
9
22
10
23
2
0
2
1
2
2
2
3
Triggering of functions (remote control)
Assignment of the input lines see section 4.4.1
Input
Decimal
Function
23
10
22
9
code
(default settings)
H
H
H
H
0
Inactive
H
H
H
L
1
PROG R/S
H
H
L
H
2
PUMP R/S
H
H
L
L
3
FILL B/STEP
H
L
H
H
4
FILL A
H
L
H
L
5
REPORT
H
L
L
H
6
ZERO
H
L
L
L
7
MARK
L
H
H
H
8
INJECT A
L
H
H
L
9
ZERO OFF
L
H
L
H
10
PLOT
L
H
L
L
11
Switch polarity
L
L
H
H
12
INJECT B
L
L
H
L
13
SELECT
L
L
L
H
14
QUIT
L
L
L
L
15
ENTER
H
L
t
P
> 100 ms
+5V
11
24
12
not assigned