Metrohm 732 IC Detector User Manual

Page 124

Advertising
background image

4.7 Programming

732 IC Detector / 733 IC Separation Center

117

4.7.4

<METHOD> key

METHOD

9

The <METHOD> key is used for the management of maximum 9 user-

defined methods, which can be stored and recalled under freely

selectable names. A method comprises the measurement parameters

defined under the <PARAM> key and the program defined under the

<PROGRAM> key.
The <METHOD> key opens the following main menu:

METHOD
>METHOD/recall

Recall method

METHOD
>METHOD/store

Store method

METHOD
>METHOD/delete

Delete method

Repeated pressing of the <EVENT> key selects the submenus in turn.

The individual inquiries of a submenu are accessed using the

<ENTER> key and exited with the <QUIT> key. The following listings

show all dialog items which appear under <METHOD>. The values

shown in the displays are the default values, the possible entry values

and ranges are shown below the display.

>METHOD
>METHOD/recall

Recall existing method

>METHOD/recall
name: DEFAULT

DEFAULT, Prep-MSM,

XXXXXXXX

Selection of the method
Use of the <SELECT> key to select the existing

method which should be recalled.

DEFAULT

Default method with empty program and

parameters set to the default values.

Prep-MSM

Method for conditioning the suppressor.

The suppressor module is switched to the

next position every 20 min.

XXXXXXXX

Name of the user method stored.

After confirmation of the name with the <ENTER> key,

the method is loaded in the working memory. The pa-

rameters under the <PARAM> and <PROGRAM>

keys will be overwritten.