Metrohm 732 IC Detector User Manual
Page 37

2 Installation
732 IC Detector / 733 IC Separation Center
30
2.7.3
Precolumns with cartridge head
The precolumn cartridges are installed in the 6.2821.040 Cartridge
head as follows (see Fig. 15):
1 Prepare separating column
•
Remove end caps from separating column 76
76.
•
Unscrew fastening screw from column inlet.
•
Take steel connector 79
79 for ferrule out of fastening screw.
2 Install cartridge
•
Remove end caps for precolumn cartridge 74
74 (the steel mesh
73
73 and gaskets 59
59 are already installed in the cartridge).
•
Mount steel spacer 78
78 on separating column 76
76.
•
Mount precolumn cartridge 74
74 on the steel spacer (comply
with flow direction if specified on the precolumn).
•
Screw Manufit pressure screw 77
77 firmly to separating column
76
76.
•
Mount steel connector 79
79 for ferrule on the inlet side of the
precolumn cartridge 74
74.
•
Screw on Manufit housing 80
80 with Manufit pressure screw 77
77.
3 Connect precolumn
•
Fit connectors to the column connection capillary 67
67
mounted on the injection valve (see section 2.5).
•
Screw column connection capillary 67
67 to Manufit housing 80
80.