2 Installation

732 IC Detector / 733 IC Separation Center

30

2.7.3

Precolumns with cartridge head

The precolumn cartridges are installed in the 6.2821.040 Cartridge

head as follows (see Fig. 15):

1 Prepare separating column

•

Remove end caps from separating column 76

76.

•

Unscrew fastening screw from column inlet.

•

Take steel connector 79

79 for ferrule out of fastening screw.

2 Install cartridge

•

Remove end caps for precolumn cartridge 74

74 (the steel mesh

73

73 and gaskets 59

59 are already installed in the cartridge).

•

Mount steel spacer 78

78 on separating column 76

76.

•

Mount precolumn cartridge 74

74 on the steel spacer (comply

with flow direction if specified on the precolumn).

•

Screw Manufit pressure screw 77

77 firmly to separating column

76

76.

•

Mount steel connector 79

79 for ferrule on the inlet side of the

precolumn cartridge 74

74.

•

Screw on Manufit housing 80

80 with Manufit pressure screw 77

77.

3 Connect precolumn

•

Fit connectors to the column connection capillary 67

67

mounted on the injection valve (see section 2.5).

•

Screw column connection capillary 67

67 to Manufit housing 80

80.