3 faults and malfunctions – Metrohm 732 IC Detector User Manual

Page 154

Advertising
background image

5.3 Faults and malfunctions

732 IC Detector / 733 IC Separation Center

147

Error message

Meaning/Causes

Rectification

range error

#

Error in internal hardware test.

Inform Metrohm service.

RS error 36

RS receive error; wrong parity.

<QUIT>, set same parity for both

devices.

RS error 37

RS receive error; wrong stop bit.

<QUIT>, set same stop bit for both

devices.

RS error 38

RS-receive error; overflow (at least 1

character could not be read).

<QUIT>, set same baud rate for both

devices, restart sender.

RS error 39

RS receive error; overflow of internal

receive buffer (>82 characters).

<QUIT>.

RS error 40

RS transmission error; DSR = OFF.

Handshake unsatisfactory for more

than 1 s.

<QUIT>, check receiver (switched on

and ready?).

RS error 41

RS transmission error; DCD = ON.

Handshake unsatisfactory for more

than 1 s.

<QUIT>, check receiver (switched on

and ready?).

RS error 42

RS transmission error; CTS = OFF.

Handshake unsatisfactory for more

than 1 s.

<QUIT>, check receiver (switched on

and ready?).

RS error 43

RS transmission error; transmission

of the IC Detector interrupted with

XOFF for at least 3 s.

<QUIT> or send XON.

RS error 44

RS transmission error; the RS232

parameters for the two devices are

no longer the same.

<QUIT>, reset RS232 parameters for

both devices.

RS error 45

RS transmission error; the receive

buffer of the IC Detector contains an

incomplete string (L

F

missing), the

transmission is thus blocked.

<QUIT> or send L

F

.

RS error 60

…

89

Error in RS232 test.

<QUIT>, check connection between the

two RS232 interfaces.

RS error XX at 709

RS error 36

…

89 at the RS232

interface "709 IC Pump".

See individual RS errors.

RS error XX at RS 232

RS error 36

…

89 at the RS232

interface "RS 232".

See individual RS errors.

sensitivity error

#

Error in internal hardware test.

Inform Metrohm service.

stopped:YY-MM-DD HH:MM

709 IC Pump stopped at specified

time owing to violation of shutoff

limits.

Press the 709 IC Pump keys <EXT.>,

<R/S> in succession and then <EXT.>

again.

store operation failed

Storage of the method failed.

Switch instrument off/on. Store method

again. If error reappears, inform

Metrohm service.

U cell error

#

Error in internal hardware test.

Inform Metrohm service.

Advertising