4.4 Basic settings
732 IC Detector / 733 IC Separation Center
91
>CONFIG/print meas.value
time interval 1.0 s
0.4
…
99999 s
Time interval for measured value printout
This inquiry appears only with "
print crit. = time
".
Time interval in seconds between printout of the indi-
vidual measured values.
>CONFIG/print meas.value
stop time off min
1
…
999 min, off
Stop time for measured value printout
This inquiry appears only with "
print crit. = time
".
Time to stop of the measured value printout.
off
No time restriction
(press <CLEAR>).
>CONFIG/print meas.value
date&time: off
on, off
Printout of date and time
This inquiry appears only with "
print crit. = immed.
" or "
crit. = time
".
on
Date and time are printed out for every
measured value.
off
Date and time are not printed out.
>CONFIG
>CONFIG/auxiliaries
General instrument settings
>CONFIG/auxiliaries
run number 0
0
…
999, off
Run number
The run number is incremented by +1 for each new
determination as follows:
without program Increase with every "INJECT A".
with program
Increase with every new start and
with every "
return
" in cycle pro-
grams.
The parameter value has the following meaning:
0
…
999
Start point for numbering.
off
No numbering
(press <CLEAR>).
>CONFIG/auxiliaries
number of cycles 1
1
…
999
Number of cycles
This parameter is identical to the parameter of the same
name under the <PROGRAM> key (see section 4.7.1)
and determines how many times cycle programs should
be run. As access to the <PROGRAM> key is blocked
while a program is running, the number of cycles can
subsequently be changed here.