Metrohm 732 IC Detector User Manual

Page 98

Advertising
background image

4.4 Basic settings

732 IC Detector / 733 IC Separation Center

91

>CONFIG/print meas.value
time interval 1.0 s

0.4

99999 s

Time interval for measured value printout

This inquiry appears only with "

print crit. = time

".

Time interval in seconds between printout of the indi-

vidual measured values.

>CONFIG/print meas.value
stop time off min

1

999 min, off

Stop time for measured value printout

This inquiry appears only with "

print crit. = time

".

Time to stop of the measured value printout.

off

No time restriction

(press <CLEAR>).

>CONFIG/print meas.value
date&time: off

on, off

Printout of date and time

This inquiry appears only with "

print crit. = immed.

" or "

print

crit. = time

".

on

Date and time are printed out for every

measured value.

off

Date and time are not printed out.

>CONFIG
>CONFIG/auxiliaries

General instrument settings

>CONFIG/auxiliaries
run number 0

0

999, off

Run number
The run number is incremented by +1 for each new

determination as follows:

without program Increase with every "INJECT A".
with program

Increase with every new start and

with every "

return

" in cycle pro-

grams.

The parameter value has the following meaning:

0

999

Start point for numbering.

off

No numbering

(press <CLEAR>).

>CONFIG/auxiliaries
number of cycles 1

1

999

Number of cycles
This parameter is identical to the parameter of the same

name under the <PROGRAM> key (see section 4.7.1)

and determines how many times cycle programs should

be run. As access to the <PROGRAM> key is blocked

while a program is running, the number of cycles can

subsequently be changed here.