6 Interfaces

732 IC Detector / 733 IC Separation Center

182

6.1.10 RS232 error rectification

Problem

Questions for corrective action

No characters can be
received on an attached
printer.

⇒

Are the devices switched on and the connecting cables
plugged in properly?

⇒

Is the printer set to ”on-line”?

⇒

Do the baud rate, data bits and parity for the two devices have
the same setting?

⇒

Is the handshake set properly?

If everything appears ok, try to print out a report by pressing the
<REPORT> key.

No data transmission
occurs and an error
message appears in the
display of the 732 IC
Detector.

⇒

⇒ RS error 36…

…39: Receive errors.

Do the RS232 data transmission parameters of the two de-
vices have the same setting?

⇒

RS error 40

…

…42: transmission errors.

Is the cable used correctly wired and plugged in?
Is the printer switched on and set to ”on-line”?

⇒

RS error 43: Data output of the 732 IC Detector blocked for
longer than 3 s by XOFF.

The received characters
are garbled.

⇒

Do the data bits and parity of the two devices have the same
setting?

⇒

Does the baud rate of the two devices have the same setting?

⇒

Is the correct printer selected?

⇒

The data transfer has been interrupted by the hardware during
a printout. Set up connections again, switch printer off then
on.

The graphics plot is not
printed out properly. Other
reports are ok.

A handshake is necessary for the graphics plot.

⇒

Is your cable correctly wired? (The DTR of the printer must be
wired with the CTS of the 732 IC Detectors.)

⇒

Set the handshake on the 732 IC Detector to "

HWs

". The

printer must be so configured that its DTR is set (usually with
DIP switches).