4 Operation

732 IC Detector / 733 IC Separation Center

108

<PROG R/S>

â

Program steps 1

k

â

return

â

Program steps m

n

â

end

No. of cycles

remote

Single program with start via remote

A remote program is started externally via a

remote input line (see section 6.2.1) and

executed once only.

inject

Single program with start via "INJECT"

An inject program is started each time injec-

tion valve A is switched to the "INJECT" posi-

tion and executed once only.

>PROGRAM/parameters
number of cycles 1

1

999

Number of cycles

This inquiry appears only with "

PROGRAM/parameters/type =

cycle

"

.

Number of executions of loop programs.

This parameter is identical to that of the same name in

the submenu

"

>CONFIG/auxiliaries

"

of the <CON-

FIG> key (see section 4.4.2). As access to the

<PROGRAM> key is blocked while a program is

running, the number of cycles can be changed only

there.

>PROGRAM/parameters
status: disabled

enabled, disabled, test

Program status
The three following program statuses are possible:

enabled

Program enabled

Irrespective of the program type, a pro-

gram can be started only when it is en-

abled. This status is shown by the lighting

up of the green LED 55 in the <PROG

R/S> key. If a method with program

steps is loaded, the program status is

automatically set to "

enabled

".

disabled

Program disabled

A disabled program can not be started. In

this status, the green LED 55 in the

<PROG R/S> key remains off. When the

instrument is switched on, the program

status is always set to

"

disabled

"

.