4 cal cond - conductivity standard – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1070

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1054

■■■■■■■■

tiamo 2.2 (ProcessLab)

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list. Max.
number of entries?

712
Range

0.4 ... 999999 s

Default value

2.0 s

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Temperature

Measurement temperature

Temperature of the solution to be measured. If a temperature sensor is con-
nected, the temperature entered here is ignored.

Range

-20.0 ... 150.0 °C

Default value

20.0

5.6.5.8.4

CAL Cond - Conductivity standard

Tab: Method

CAL Cond Properties... Conductivity standard

Command name

Name of the command.

Input

25 characters

Definition of the conductivity standard
Selection of the type as to how the data are to be defined for the conductivity
standard.

Advertising