5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

719

Titrando
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

U(pol)

The polarization voltage is the potential applied to the polarized electrode
during an amperometric measurement.

Titrando, 855
Range

-1250 ... 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Range

-1270 ... 1270 mV (Increment: 10)

Default value

400 mV

Electrode test

on | off (Default value: off)

If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.